ARES MB FLAT PANEL : INVERTER KV ERROR -- a) High voltage circuit in the monobloc is unbalanced or kV feed-back is not present during an exposure or it is out of range. The yellow B11-Ld1 led is ON. b) During x-ray emission, kV decrease under 75% in respect of the set value (during the emission the led B11-Ld5 turns OFF). c) During X-ray emission, kV increase over 110% of set value (software control on TUBE KV analog signal). -- Perform an exposure and check on the Inverter B11 board the following: - KV FEED signal (1V = 20kV) - the KV+ and kV- signals must be balanced - kV>75% signal - TUBE KV signal (1V = 20kV) --- B11-Tp1 B11-Tp5 and Tp6 B11-Ld5ON B2-Tp6 -------------------------------- INVERTER OVERLOAD -- The current absorbed by the Inverter is too high. The red B11-Ld2 led is ON (Ipk>300A). Monobloc fault or too high anodical current. -- In preparation phase check on B7 board the following signals: - FIL DAC OUT - mA SET - FIL CUR in radiography phase - ANODIC mA - mA STABILIZATION The replacement of the monobloc could be necessary. -- B7-Tp3 B7-Tp1 (1V=20mA) B7-Tp8 (1V=100mA) B11-Tp8 (1V=20mA) B2-Ld26 ------------------------------- INVERTER FAULT -- Error in the driver of the IGBTs. The red B11-Ld3-4 led is ON. -- Perform an exposure. Replace the Inverter if the fault persists. -----------------------------