-----------------General Fluoro Setup---------------- Fluoro Timer Mode: 5 min Min Fluoro kV: 40 Max Fluoro kV: 120 Max II Modes: 0 Fluoro ABS: Enabled Fluoro ABS Default: None Fluoro/Rad kV Def.: None HLF Select: Console PF Select: Console ---------Fluoro to Rad kV Transfer Curves------------ F/R kV Curve1: Enabled F/R kV Curve2: Disabled F/R kV Curve3: Disabled F/R 40kV Curve1: 40 F/R 45kV Curve1: 45 F/R 50kV Curve1: 50 F/R 55kV Curve1: 55 F/R 60kV Curve1: 60 F/R 65kV Curve1: 65 F/R 70kV Curve1: 70 F/R 75kV Curve1: 75 F/R 80kV Curve1: 80 F/R 85kV Curve1: 85 F/R 90kV Curve1: 90 F/R 95kV Curve1: 95 F/R 100kV Curve1: 100 F/R 105kV Curve1: 105 F/R 110kV Curve1: 110 F/R 115kV Curve1: 115 F/R 120kV Curve1: 120 F/R 125kV Curve1: 125 F/R 40kV Curve2: 50 F/R 45kV Curve2: 51 F/R 50kV Curve2: 53 F/R 55kV Curve2: 55 F/R 60kV Curve2: 57 F/R 65kV Curve2: 59 F/R 70kV Curve2: 61 F/R 75kV Curve2: 64 F/R 80kV Curve2: 66 F/R 85kV Curve2: 69 F/R 90kV Curve2: 72 F/R 95kV Curve2: 74 F/R 100kV Curve2: 76 F/R 105kV Curve2: 79 F/R 110kV Curve2: 81 F/R 115kV Curve2: 82 F/R 120kV Curve2: 83 F/R 125kV Curve2: 84 F/R 40kV Curve3: 55 F/R 45kV Curve3: 58 F/R 50kV Curve3: 60 F/R 55kV Curve3: 63 F/R 60kV Curve3: 65 F/R 65kV Curve3: 68 F/R 70kV Curve3: 70 F/R 75kV Curve3: 75 F/R 80kV Curve3: 75 F/R 85kV Curve3: 77 F/R 90kV Curve3: 80 F/R 95kV Curve3: 83 F/R 100kV Curve3: 85 F/R 105kV Curve3: 87 F/R 110kV Curve3: 90 F/R 115kV Curve3: 93 F/R 120kV Curve3: 95 F/R 125kV Curve3: 98 -----------------Continuous Fluoro ------------------ Fluoro ABS Level 0: 5 Fluoro ABS Level 1: 5 Fluoro ABS Level 2: 5 Fluoro ABS Level 3: 5 Fluoro ABS Gain: 10 Fluoro ABS Deadband: 3 Fluoro ABS Delay: 150 Fluoro ABS Initial kV: 75 Fluoro Signal Gain: 10 Fluoro ABS Curve Def.: Curve 1 Fluoro ABS Curve1: Enabled Fluoro ABS Curve2: Fluoro ABS Curve3: Disabled Enabled Fluoro Curve 1: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 0.5 1 2 3 4 5 6 7 8 8 Fluoro Curve 2: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 0.5 0.5 0.5 0.5 1.3 3 4 4.2 4.9 4.5 Fluoro Curve 3: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 0.5 0.5 1.3 3 4 6 5.5 4.4 3.7 3.5 Fluoro Man Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 0.5 1 1.6 2.1 2.7 3.2 3.8 4.4 5 5 Fluoro Man Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 8 8 8 8 7 5.8 4.9 4.4 4 3.8 Fluoro ABS Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 5 5 5 5 5 5 5 5 5 5 Fluoro ABS Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 10 10 10 10 10 10 9.5 8.4 7.7 7.6 Fluoro Mag 1 mA Mult: 1.2 Fluoro Mag 2 mA Mult: 1.7 Fluoro Mag 3 mA Mult: 2.1 -------------------High Level Fluoro----------------- HLF Mode: Disabled HLF mA Factor: 3 HLF ABS Level 0: 2 HLF ABS Level 1: 2.13 HLF ABS Level 2: 2.61 HLF ABS Level 3: 3.53 HLF Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 15 15 15 15 14 9 5 3.3 2 1.5 HLF Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 20 20 20 20 19 14 10 8.3 7 6.5 HLF Mag 1 mA Multiplier: 1.1 HLF Mag 2 mA Multiplier: 1.6 HLF Mag 3 mA Multiplier: 2.1 ------------------Pulsed Fluoro Setup---------------- PF Mode: Enabled PF Default: None PPS Default: None PF Default mA: 40 PF Default ms: 8 PF ABS Curve Def.: None PF Sync. Frequency: 50 Hz PF ABS Level 0: 5 PF ABS Level 1: 5 PF ABS Level 2: 5 PF ABS Level 3: 5 PF ABS Gain: 8 PF ABS Deadband: 10 PF ABS Delay: 150 PF ABS Initial kV: 75 PF Signal Gain: 10 PF Integration: Disabled PF ABS Curve1: Disabled PF ABS Curve2: Disabled PF ABS Curve3: Disabled PF mA Curve 1: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 5 5 5 5 5 5 8 12 20 20 PF mA Curve 2: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 5 5 12 12 17 20 25 25 25 25 PF mA Curve 3: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 10 10 20 25 25 25 25 30 35 35 PF ms Curve 1: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 2 2 2 2 2 4 5 7 7 8 PF ms Curve 2: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 3 4 4 4 5 5 8 8 8 9 PF ms Curve 3: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 3 6 6 7 11 13 13 13 16 16 PF mA Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 30 30 30 30 30 30 27 22 16 16 PF mA Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 30 30 30 30 30 30 27 22 16 16 PF ms Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 16 16 16 16 13 9 8 8 8 8 PF ms Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 16 16 16 16 13 9 8 8 8 8 PF 1PPS ABS Comp.: 0 PF 3PPS ABS Comp.: 0 PF 7.5PPS ABS Comp.: 0 PF 15PPS ABS Comp.: 0 PF 30PPS ABS Comp.: 0 PF 1PPS: Enabled PF 3PPS: Enabled PF 7.5PPS: Enabled PF 15PPS: Enabled PF 30PPS: Enabled PF Mag 1 ms Multiplier: 1.2 PF Mag 2 ms Multiplier: 2 PF Mag 3 ms Multiplier: 2.6 -----------High Level Pulsed Fluoro Setup------------- HPF Mode: Disabled HPF mA Factor: 25 HPF ms Factor: 2 HPF ABS Level 0: 2.3 HPF ABS Level 1: 2 HPF ABS Level 2: 2.6 HPF ABS Level 3: 4.5 HPF mA Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 60 60 60 60 60 60 50 40 35 35 HPF mA Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 60 60 60 60 60 60 50 40 35 35 HPF ms Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 16 16 16 16 13 10 10 10 10 10 HPF ms Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 16 16 16 16 13 10 10 10 10 10 PF 1PPS: Enabled PF 3PPS: Enabled PF 7.5PPS: Enabled PF 15PPS: Enabled PF 30PPS: Enabled HPF Mag 1 ms Multiplier: 1.2 HPF Mag 2 ms Multiplier: 2 HPF Mag 3 ms Multiplier: 2.4