Chapter 8 SERVICE This chapter explain several function of instrument, those function can help user operate the instrument conveniently and amusive. At the blood cell analyzer windows, press the right key of the mouse, choose the Function/Service. 8.1 FLUSH This fuction is use for solve the usual probe clog. There will be a fixed pressure and voltage effect to the probe to get rid of the clog. 8.2 ENHANCED FLUSH This function is use to solve stubborn probe clog. Probe cleaner would be add in WBC and RBC count pool, the probe dip in the cleaner to get rid of the stubborn clog. 8.3 RINSE This function use to clean probe, sample, measure tube routinely. 8.4 ENHANCED RINSE This function use to clean probe, sample, measure tube enhancive. 8.5 DRAIN CHAMBERS This function is used for observe the count pool and get rid of the liquid in count pool. 8.6 PARK If the instrument is not going to be used in 2 week or longer, use this function to clean the instrument. Prepare distilled water. 8.7 PRIME DILUTION Fill dilution to relevant tube. 8.8 PRIME LYSE Fill lyse to relevant tube. 8.9 PRIME RINSE Fill Rinse to relevant tube. 8.10 MAINTENANCE User can finish the weekly maintenance follow the screen prompt. Prepare rinse and probe cleaner first. 8.11 MECHANIC The password is :6100. The instrument turn into detection windows, valve, motor, circuit and connector would be detected to affirm the malfunction. Valve: "ON" means the valve is on. "OFF" means the valve is off Motor: "OK" means the motor is in the course of nature. "FAIL" means the motor is failure. Circuit: "OK" means the circuit is in the course of nature. "FAIL"means the circuit is failure. "OFF" means no negative pressure. "ON" means have negative pressure. "LOW" means the temperature is lower than 18°C Function of menu and shortcut key: Mute: Close the alarm voice. Help: turn into help windows, obtain the help on line. Back: Quit currently windows, back to blood cell analyzer windows. 8.12 SPECIAL The instrument turn into special maintenance windows, and it could adjust the control data. This function should operated by the guide of engineers of SFRI.