Generator Installation Report Date (dd/mm/yy): 7/2/2007 Power Revision: 4.22 --------------------Tube Data-------------------- Tube Number: 1 Tube Revision: 1.1 Tube Angle: 12 deg. Tube Speed: low Tube Frequency Rating: 60 Hz Tube Name: RTM90HS 0.6/1.2 Low Speed maximum SF power at 100 ms: 20 kW Low Speed maximum LF power at 100 ms: 44.1 kW Low Speed maximum MF power at 100 ms: --- High Speed maximum SF power at 100 ms: 0 kW High Speed maximum LF power at 100 ms: 0 kW High Speed maximum MF power at 100 ms: --- Filament Boost Time: 200 ms Filament Preheat Time: 800 ms Micro Filament Standby Current: --- Micro Filament Maximum Current: --- Small Filament Standby Current: 2.1 A Small Filament Maximum Current: 5.2 A Large Filament Standby Current: 2.1 A Large Filament Maximum Current: 5.4 A Anode Warning: 80 % Anode Limit: 90 % Maximum Tube Voltage: 150 kV Maximum Small Filament Current: 160 mA Maximum Micro Filament Current: --- --------------------Tube Data-------------------- Tube Number: 2 Tube Revision: 1.2 Tube Angle: 12 deg. Tube Speed: dual Tube Frequency Rating: 60 Hz Tube Name: A292 0.6/1.2 Low Speed maximum SF power at 100 ms: 25 kW Low Speed maximum LF power at 100 ms: 62.5 kW Low Speed maximum MF power at 100 ms: --- High Speed maximum SF power at 100 ms: 39.6 kW High Speed maximum LF power at 100 ms: 96 kW High Speed maximum MF power at 100 ms: --- Filament Boost Time: 200 ms Filament Preheat Time: 800 ms Micro Filament Standby Current: --- Micro Filament Maximum Current: --- Small Filament Standby Current: 2.3 A Small Filament Maximum Current: 5.3 A Large Filament Standby Current: 2.6 A Large Filament Maximum Current: 5.3 A Anode Warning: 80 % Anode Limit: 90 % Maximum Tube Voltage: 150 kV Maximum Small Filament Current: 320 mA Maximum Micro Filament Current: --- --------------------Generator Limits-------------------- Maximum Generator Power: 65 kW Maximum Generator Current: 800 mA Minimum Generator Current: 10 mA Maximum Generator Current-Time: 630 mAs --------------------Receptor Setup----------------------- Receptor Number: 1 2 3 4 5 6 Tube Number: 1 1 1 1 0 0 Tomo Enabled: on off off on off off Fluoro Enabled: on off off off on on Serial Enabled: on off off off off off AEC Channel: 1 off 2 off 1 1 Interface Options: none none none none none none Fluoro Hangover (s): 30 30 30 30 30 30 Rad Hangover (s): 0 0 0 0 0 0 Last Image Hold (ms): 40 40 40 40 40 40 Memory: off off off dflt off off Tomo Back-Up Time (ms): 5000 2500 2500 5000 2500 2500 Auto Focus: off off off off off off AEC Back-Up Mode: fixed fixed fixed fixed fixed fixed AEC Back-Up mAs: 500 500 500 500 500 500 AEC Back-Up ms: 3200 3200 3200 3200 3200 3200 AEC Lock Exp: 3 3 3 3 3 3 Default Voltage (kV): 64 75 75 50 75 75 Default Current (mA): 200 200 200 100 200 200 Default Time (ms): 4000 200 200 4000 200 200 Technique: mA/ms mAs mAs mA/ms mAs mAs Default Focus: small large large large large large Default Density: 0 0 0 0 0 0 Default Film Screen: II I I I I I Default Fields: RL RCL RCL RCL RCL RCL Functional Options: none none none none none none --------------------Receptor I/O-------------------- Receptor: 1 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off on on off Bucky 2 Select: off off off off off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off off off Spare: on on on on on Tomo/Bucky Start: off off off off off ALE: off off off on on Collimator Bypass: off off off off off Room Light: off on on on on (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off on off Remote Prep: off on off off off Rem. Fluoro Exp.: off off off off on Console Exposure: off off off off off Console Prep: off off off off off Tomo Exposure: off off off on off Remote Tomo Select: on off off off off Receptor: 2 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off off off off Bucky 2 Select: off off off off off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off off off Spare: off off off off off Tomo/Bucky Start: off off off off off ALE: off off off off off Collimator Bypass: off off off off off Room Light: off on on on on (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off off off Remote Prep: off off off off off Rem. Fluoro Exp.: off off off off off Console Exposure: off off off on off Console Prep: off on off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off Receptor: 3 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off off off off Bucky 2 Select: off off off off off Bucky 3 Select: off off off on off Tomo/Bucky 4 Sel.: off off off off off Spare: off off off off off Tomo/Bucky Start: off off off off off ALE: off off off off off Collimator Bypass: off off off off off Room Light: off on on on on (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off on off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off off off Remote Prep: off off off off off Rem. Fluoro Exp.: off off off off off Console Exposure: off off off on off Console Prep: off on off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off Receptor: 4 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off on on off Bucky 2 Select: off off off off off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off off off Spare: on on on on on Tomo/Bucky Start: off off off off off ALE: off off off on on Collimator Bypass: off off off off off Room Light: off on on on on (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off on off Remote Prep: off on off off off Rem. Fluoro Exp.: off off off off off Console Exposure: off off off off off Console Prep: off off off off off Tomo Exposure: off off off on off Remote Tomo Select: on off off off off Receptor: 5 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off off on off Bucky 2 Select: off off off on off Bucky 3 Select: off off off on off Tomo/Bucky 4 Sel.: off off off off off Spare: off off off off off Tomo/Bucky Start: off off off on off ALE: off off off on off Collimator Bypass: off off off on off Room Light: off off off off off (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off off off Remote Prep: off off off off off Rem. Fluoro Exp.: off off off off on Console Exposure: off off off on off Console Prep: off on off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off Receptor: 6 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off off off off Bucky 2 Select: off off off off off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off on off Spare: off off off off off Tomo/Bucky Start: off off off on off ALE: off off off on off Collimator Bypass: off off off on off Room Light: off off off off off (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off off off Remote Prep: off off off off off Rem. Fluoro Exp.: off off off off on Console Exposure: off off off on off Console Prep: off on off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off --------------------AEC Setup-------------------- Channel: 1 Media: Film Film Screen Enable: I, II, III Fields: right centre left Chamber Type: Solid State Right Field Compensation: 0 Center Field Compensation: 0 Left Field Compensation: 0 Film Screen 1 Gain: 10 Film Screen 2 Gain: 10 Film Screen 3 Gain: 10 Channel: 2 Media: Film Film Screen Enable: I, II, III Fields: right centre left Chamber Type: Solid State Right Field Compensation: 0 Center Field Compensation: 0 Left Field Compensation: 0 Film Screen 1 Gain: 10 Film Screen 2 Gain: 10 Film Screen 3 Gain: 10 Channel: 3 Media: Film Film Screen Enable: I, II, III Fields: right centre left Chamber Type: Solid State Right Field Compensation: 0 Center Field Compensation: 0 Left Field Compensation: 0 Film Screen 1 Gain: 10 Film Screen 2 Gain: 10 Film Screen 3 Gain: 10 Channel: 4 Media: Film Film Screen Enable: I, II, III Fields: right centre left Chamber Type: Solid State Right Field Compensation: 0 Center Field Compensation: 0 Left Field Compensation: 0 Film Screen 1 Gain: 10 Film Screen 2 Gain: 10 Film Screen 3 Gain: 10 --------------------AEC Calibration Film--------------- Film Screen: 1 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 8.40 7.80 6.60 5.40 4.28 3.80 3.40 2.80 RLF Compensation at 50 ms: 0 RLF Compensation at 500 ms: 0 RLF Compensation at 1000 ms: 0 Multiple Spot Compensation: 0 Film Screen: 2 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 6.80 6.40 5.60 4.80 4.28 3.80 3.40 3.00 RLF Compensation at 50 ms: 0 RLF Compensation at 500 ms: 0 RLF Compensation at 1000 ms: 0 Multiple Spot Compensation: 0 Film Screen: 3 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 5.40 5.20 4.80 4.48 4.20 3.80 3.40 2.80 RLF Compensation at 50 ms: 0 RLF Compensation at 500 ms: 0 RLF Compensation at 1000 ms: 0 Multiple Spot Compensation: 0 --------------------AEC Calibration Digital------------ Dgital Rad Curve: 1 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 50.0 50.0 50.0 50.0 50.0 50.0 50.0 50.0 Dgital Rad Curve: 2 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 50.0 50.0 50.0 50.0 50.0 50.0 50.0 50.0 Dgital Rad Curve: 3 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 50.0 50.0 50.0 50.0 50.0 50.0 50.0 50.0 --------------------AEC Densities-------------------- -8 Density: 0 -7 Density: 0 -6 Density: 0 -5 Density: 62 -4 Density: 50 -3 Density: 37 -2 Density: 25 -1 Density: 12 1 Density: 12 2 Density: 25 3 Density: 37 4 Density: 50 5 Density: 62 6 Density: 0 7 Density: 0 8 Density: 0 -----------------General Fluoro Setup---------------- Fluoro Timer Mode: 5 min Min Fluoro kV: 40 Max Fluoro kV: 125 Max II Modes: 0 Fluoro ABS: Enabled Fluoro ABS Default: On Fluoro/Rad kV Def.: None HLF Select: Console PF Select: Console ---------Fluoro to Rad kV Transfer Curves------------ F/R kV Curve1: Enabled F/R kV Curve2: Disabled F/R kV Curve3: Disabled F/R 40kV Curve1: 40 F/R 45kV Curve1: 45 F/R 50kV Curve1: 50 F/R 55kV Curve1: 55 F/R 60kV Curve1: 60 F/R 65kV Curve1: 65 F/R 70kV Curve1: 70 F/R 75kV Curve1: 75 F/R 80kV Curve1: 80 F/R 85kV Curve1: 85 F/R 90kV Curve1: 90 F/R 95kV Curve1: 95 F/R 100kV Curve1: 100 F/R 105kV Curve1: 105 F/R 110kV Curve1: 110 F/R 115kV Curve1: 115 F/R 120kV Curve1: 120 F/R 125kV Curve1: 125 F/R 40kV Curve2: 50 F/R 45kV Curve2: 51 F/R 50kV Curve2: 53 F/R 55kV Curve2: 55 F/R 60kV Curve2: 57 F/R 65kV Curve2: 59 F/R 70kV Curve2: 61 F/R 75kV Curve2: 64 F/R 80kV Curve2: 66 F/R 85kV Curve2: 69 F/R 90kV Curve2: 72 F/R 95kV Curve2: 74 F/R 100kV Curve2: 76 F/R 105kV Curve2: 79 F/R 110kV Curve2: 81 F/R 115kV Curve2: 82 F/R 120kV Curve2: 83 F/R 125kV Curve2: 84 F/R 40kV Curve3: 55 F/R 45kV Curve3: 58 F/R 50kV Curve3: 60 F/R 55kV Curve3: 63 F/R 60kV Curve3: 65 F/R 65kV Curve3: 68 F/R 70kV Curve3: 70 F/R 75kV Curve3: 75 F/R 80kV Curve3: 75 F/R 85kV Curve3: 77 F/R 90kV Curve3: 80 F/R 95kV Curve3: 83 F/R 100kV Curve3: 85 F/R 105kV Curve3: 87 F/R 110kV Curve3: 90 F/R 115kV Curve3: 93 F/R 120kV Curve3: 95 F/R 125kV Curve3: 98 -----------------Continuous Fluoro ------------------ Fluoro ABS Level 0: 1 Fluoro ABS Level 1: 0.6 Fluoro ABS Level 2: 1.1 Fluoro ABS Level 3: 1.6 Fluoro ABS Gain: 20 Fluoro ABS Deadband: 3 Fluoro ABS Delay: 30 Fluoro ABS Initial kV: 75 Fluoro Signal Gain: 10 Fluoro ABS Curve Def.: None Fluoro ABS Curve1: Enabled Fluoro ABS Curve2: Fluoro ABS Curve3: Disabled Enabled Fluoro Curve 1: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 0.5 0.5 0.5 0.5 0.5 0.7 1.5 2.5 2.9 3 Fluoro Curve 2: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 0.5 0.5 0.5 0.5 1.3 3 4 4.2 4.9 4.5 Fluoro Curve 3: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 0.5 0.5 1.3 3 4 6 5.5 4.4 3.7 3.5 Fluoro Man Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 6 6 6 6 5 3.8 2.9 2.4 2 1.8 Fluoro Man Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 8 8 8 8 7 5.8 4.9 4.4 4 3.8 Fluoro ABS Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 6 6 6 6 6 6 5.5 4.4 3.7 3.6 Fluoro ABS Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 10 10 10 10 10 10 9.5 8.4 7.7 7.6 Fluoro Mag 1 mA Mult: 1.2 Fluoro Mag 2 mA Mult: 1.7 Fluoro Mag 3 mA Mult: 2.1 -------------------High Level Fluoro----------------- HLF Mode: Disabled HLF mA Factor: 3 HLF ABS Level 0: 2 HLF ABS Level 1: 2.13 HLF ABS Level 2: 2.61 HLF ABS Level 3: 3.53 HLF Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 15 15 15 15 14 9 5 3.3 2 1.5 HLF Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 20 20 20 20 19 14 10 8.3 7 6.5 HLF Mag 1 mA Multiplier: 1.1 HLF Mag 2 mA Multiplier: 1.6 HLF Mag 3 mA Multiplier: 2.1 ------------------Pulsed Fluoro Setup---------------- PF Mode: Disabled PF Default: None PPS Default: None PF Default mA: 20 PF Default ms: 8 PF ABS Curve Def.: None PF Sync. Frequency: 60 Hz PF ABS Level 0: 1 PF ABS Level 1: 0.7 PF ABS Level 2: 1.2 PF ABS Level 3: 1.9 PF ABS Gain: 20 PF ABS Deadband: 5 PF ABS Delay: 30 PF ABS Initial kV: 75 PF Signal Gain: 20 PF Integration: Enabled PF ABS Curve1: Enabled PF ABS Curve2: Disabled PF ABS Curve3: Disabled PF mA Curve 1: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 5 5 5 5 5 5 8 12 20 20 PF mA Curve 2: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 5 5 12 12 17 20 25 25 25 25 PF mA Curve 3: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 10 10 20 25 25 25 25 30 35 35 PF ms Curve 1: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 2 2 2 2 2 4 5 7 7 8 PF ms Curve 2: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 3 4 4 4 5 5 8 8 8 9 PF ms Curve 3: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 3 6 6 7 11 13 13 13 16 16 PF mA Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 30 30 30 30 30 30 27 22 16 16 PF mA Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 30 30 30 30 30 30 27 22 16 16 PF ms Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 16 16 16 16 13 9 8 8 8 8 PF ms Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 16 16 16 16 13 9 8 8 8 8 PF 1PPS ABS Comp.: 0 PF 3PPS ABS Comp.: 0 PF 7.5PPS ABS Comp.: 0 PF 15PPS ABS Comp.: 0 PF 30PPS ABS Comp.: 0 PF 1PPS: Enabled PF 3PPS: Enabled PF 7.5PPS: Enabled PF 15PPS: Enabled PF 30PPS: Enabled PF Mag 1 ms Multiplier: 1.2 PF Mag 2 ms Multiplier: 2 PF Mag 3 ms Multiplier: 2.6 -----------High Level Pulsed Fluoro Setup------------- HPF Mode: Disabled HPF mA Factor: 25 HPF ms Factor: 2 HPF ABS Level 0: 2.3 HPF ABS Level 1: 2 HPF ABS Level 2: 2.6 HPF ABS Level 3: 4.5 HPF mA Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 60 60 60 60 60 60 50 40 35 35 HPF mA Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 60 60 60 60 60 60 50 40 35 35 HPF ms Min SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 16 16 16 16 13 10 10 10 10 10 HPF ms Max SID: Ref.: 40kV 50kV 60kV 70kV 80kV 90kV 100kV 110kV 120kV 125kV Val.: 16 16 16 16 13 10 10 10 10 10 PF 1PPS: Enabled PF 3PPS: Enabled PF 7.5PPS: Enabled PF 15PPS: Enabled PF 30PPS: Enabled HPF Mag 1 ms Multiplier: 1.2 HPF Mag 2 ms Multiplier: 2 HPF Mag 3 ms Multiplier: 2.4 -------------------SID Compensation------------------ SID Compensation: Disabled SID Interface Type: 0 SID Min Reference: 120 SID Max Reference: 180 --------------------DAP Setup-------------------- DAP : off DAP Mode: summation DAP Device Type 1: 0 DAP Calibration Value 1: 1 DAP Test Value 1: 250 DAP Device Type 2: 0 DAP Calibration Value 2: 1 DAP Test Value 2: 250 DAP Accumulated Alarm 0 mGycm² DAP Rate Alarm 0 mGycm²/s DAP Printer off DAP Rate Display off 27253