Generator Installation Report Date (dd/mm/yy): 21/9/2006 Power Revision: 2.68 --------------------Tube Data-------------------- Tube Number: 1 Tube Revision: 1.1 Tube Angle: 13 deg. Tube Speed: low Tube Frequency Rating: 60 Hz Tube Name: RTM101HS .6/1.3 Low Speed maximum SF power at 100 ms: 17.4 kW Low Speed maximum LF power at 100 ms: 60 kW Low Speed maximum MF power at 100 ms: --- High Speed maximum SF power at 100 ms: 30 kW High Speed maximum LF power at 100 ms: 104.8 kW High Speed maximum MF power at 100 ms: --- Filament Boost Time: 200 ms Filament Preheat Time: 800 ms Micro Filament Standby Current: --- Micro Filament Maximum Current: --- Small Filament Standby Current: 2.2 A Small Filament Maximum Current: 5.4 A Large Filament Standby Current: 2.2 A Large Filament Maximum Current: 5.4 A Anode Warning: 80 % Anode Limit: 90 % Maximum Tube Voltage: 150 kV Maximum Small Filament Current: 320 mA Maximum Micro Filament Current: --- --------------------Tube Data-------------------- Tube Number: 2 Tube Revision: 1.1 Tube Angle: 13 deg. Tube Speed: low Tube Frequency Rating: 60 Hz Tube Name: RTM101HS .6/1.3 Low Speed maximum SF power at 100 ms: 17.4 kW Low Speed maximum LF power at 100 ms: 60 kW Low Speed maximum MF power at 100 ms: --- High Speed maximum SF power at 100 ms: 30 kW High Speed maximum LF power at 100 ms: 104.8 kW High Speed maximum MF power at 100 ms: --- Filament Boost Time: 200 ms Filament Preheat Time: 800 ms Micro Filament Standby Current: --- Micro Filament Maximum Current: --- Small Filament Standby Current: 2.2 A Small Filament Maximum Current: 5.4 A Large Filament Standby Current: 2.2 A Large Filament Maximum Current: 5.4 A Anode Warning: 80 % Anode Limit: 90 % Maximum Tube Voltage: 150 kV Maximum Small Filament Current: 320 mA Maximum Micro Filament Current: --- --------------------Generator Limits-------------------- Maximum Generator Power: 50 kW Maximum Generator Current: 630 mA Minimum Generator Current: 10 mA Maximum Generator Current-Time: 630 mAs --------------------Receptor Setup----------------------- Receptor Number: 1 2 3 4 5 6 Tube Number: 1 1 0 1 0 0 Tomo Enabled: off off off on off off Fluoro Enabled: on off on on on on Serial Enabled: on on off on off off AEC Channel: 1 off 1 off 1 1 Interface Options: none none none none none none Fluoro Hangover (s): 30 30 30 30 30 30 Rad Hangover (s): 10 0 0 0 0 0 Last Image Hold (ms): 40 40 40 40 40 40 Memory: dflt dflt off dflt off off Tomo Back-Up Time (ms): 4000 4000 2500 4000 2500 2500 Auto Focus: on on off on off off AEC Back-Up Mode: fixed fixed fixed fixed fixed fixed AEC Back-Up mAs: 500 500 500 500 500 500 AEC Back-Up ms: 3200 3200 3200 3200 3200 3200 AEC Lock Exp: N/A N/A N/A N/A N/A N/A Default Voltage (kV): 75 75 75 75 75 75 Default Current (mA): 400 200 200 50 200 200 Default Time (ms): 200 200 200 4000 200 200 Technique: AEC mAs mAs mA/ms mAs mAs Default Focus: large large large small large large Default Density: 0 0 0 0 0 0 Default Film Screen: II I I I I I Default Fields: C RCL RCL RCL RCL RCL Functional Options: none none none none none none --------------------Receptor I/O-------------------- Receptor: 1 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off on on off Bucky 2 Select: off off off off off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off off off Spare: off off off off off Tomo/Bucky Start: off off off off off ALE: off off off off off Collimator Bypass: off off off off off Room Light: off off off off off (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off on off Remote Prep: off on off off off Rem. Fluoro Exp.: off off off off on Console Exposure: off off off off off Console Prep: off off off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off Receptor: 2 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off off off off Bucky 2 Select: on on on on off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off off off Spare: off off off off off Tomo/Bucky Start: off off off off off ALE: off off off off off Collimator Bypass: off off off off off Room Light: off off off off off (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off off off Remote Prep: off off off off off Rem. Fluoro Exp.: off off off off off Console Exposure: off off off on off Console Prep: off on off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off Receptor: 3 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off off off off Bucky 2 Select: off off off on off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off off off Spare: off off off off off Tomo/Bucky Start: off off off off off ALE: off off off on off Collimator Bypass: off off off off off Room Light: off off off off off (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off off off Remote Prep: off off off off off Rem. Fluoro Exp.: off off off off on Console Exposure: off off off on off Console Prep: off on off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off Receptor: 4 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off on on off Bucky 2 Select: off off off off off Bucky 3 Select: on on on on on Tomo/Bucky 4 Sel.: off off off off off Spare: off off off off off Tomo/Bucky Start: off off off off off ALE: off off off off off Collimator Bypass: off off off off off Room Light: off off off off off (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off on off Remote Prep: off on off off off Rem. Fluoro Exp.: off off off off on Console Exposure: off off off off off Console Prep: off off off off off Tomo Exposure: off off off on off Remote Tomo Select: on off off off off Receptor: 5 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off off off off Bucky 2 Select: off off off off off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off off off Spare: off off off off off Tomo/Bucky Start: off off off off off ALE: off off off off off Collimator Bypass: off off off off off Room Light: off off off off off (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off off off Remote Prep: off off off off off Rem. Fluoro Exp.: off off off off on Console Exposure: off off off on off Console Prep: off on off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off Receptor: 6 State: Standby Prep Gen Ready Rad Exp Fluoro Exp (Outputs) Bucky 1 Select: off off off off off Bucky 2 Select: off off off off off Bucky 3 Select: off off off off off Tomo/Bucky 4 Sel.: off off off on off Spare: off off off off off Tomo/Bucky Start: off off off off off ALE: off off off on off Collimator Bypass: off off off off off Room Light: off off off off off (Inputs) I.I Safety: off off off off off Collimator Interlock: off off off off off Bucky Contacts: off off off off off Spare: off off off off off Thermal Switch 1: on on off on on Thermal Switch 2: off off off off off Room Door Intlk: off on off on on Multiple Spot Exposure: off off off off off Remote Exposure: off off off off off Remote Prep: off off off off off Rem. Fluoro Exp.: off off off off on Console Exposure: off off off on off Console Prep: off on off off off Tomo Exposure: off off off off off Remote Tomo Select: off off off off off --------------------AEC Setup-------------------- Channel: 1 Film Screen Enable: I, II, III Fields: right centre left Chamber Type: Apelem Right Field Compensation: 0 Centre Field Compensation: 0 Left Field Compensation: 0 Channel: 2 Film Screen Enable: I, II, III Fields: right centre left Chamber Type: Apelem Right Field Compensation: 0 Centre Field Compensation: 0 Left Field Compensation: 0 Channel: 3 Film Screen Enable: I, II, III Fields: right centre left Chamber Type: Apelem Right Field Compensation: 0 Centre Field Compensation: 0 Left Field Compensation: 0 Channel: 4 Film Screen Enable: I, II, III Fields: right centre left Chamber Type: Apelem Right Field Compensation: 0 Centre Field Compensation: 0 Left Field Compensation: 0 --------------------AEC Calibration Film--------------- Film Screen: 1 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 8.40 7.80 6.60 5.40 4.28 3.80 3.40 2.80 RLF Compensation at 50 ms: 0 RLF Compensation at 500 ms: 0 RLF Compensation at 1000 ms: 0 Multiple Spot Compensation: 0 Film Screen: 2 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 6.80 6.40 5.60 4.80 4.28 3.80 3.40 3.00 RLF Compensation at 50 ms: 0 RLF Compensation at 500 ms: 0 RLF Compensation at 1000 ms: 0 Multiple Spot Compensation: 0 Film Screen: 3 References: 50 kV 55 kV 65 kV 75 kV 85 kV 95 kV 110 kV 130 kV 5.40 5.20 4.80 4.48 4.20 3.80 3.40 2.80 RLF Compensation at 50 ms: 0 RLF Compensation at 500 ms: 0 RLF Compensation at 1000 ms: 0 Multiple Spot Compensation: 0 --------------------AEC Densities-------------------- -8 Density: 0 -7 Density: 0 -6 Density: 0 -5 Density: 62 -4 Density: 50 -3 Density: 37 -2 Density: 25 -1 Density: 12 1 Density: 12 2 Density: 25 3 Density: 37 4 Density: 50 5 Density: 62 6 Density: 0 7 Density: 0 8 Density: 0 --------------------Fluoro Setup-------------------- min Fluoro kV: 40 II Modes: 2 Fluoro-Rad kV Transfer: disabled ABS Dose 1.0 mA: 125 kV ABS Dose 2.0 mA: 125 kV ABS Dose 3.0 mA: 125 kV ABS Dose 4.0 mA: 125 kV ABS Dose 5.0 mA: 125 kV ABS Dose 6.0 mA: 125 kV Manual Dose 1.0 mA: 125 kV Manual Dose 2.0 mA: 125 kV Manual Dose 3.0 mA: 125 kV Manual Dose 4.0 mA: 125 kV Manual Dose 5.0 mA: 125 kV Manual Dose 6.0 mA: 125 kV Loop Gain: 6 Nominal Dose: 140 Dose Level 1: 175 Dose Level 2: 210 Auto mA/kV Curve: 3 ABS Channel: external channel ABS Default: none High Level Fluoro: off HLF Default: off HLF ABS Range: off HLF mA/kV Curve Limit: 3 HLF mA/kV Curve: kV only HLF ABS Default: no default FL Timer Mode: 5 min HLF mA Limit 20.0 PF On: off PF mA: 20 PF ms: 6 PF Default: no default ABS Signal Gain: 13 PF ABS Signal Gain: 20 PF 3PPS Dose Limit: 125 PF 7.5PPS Dose Limit: 125 PF 15PPS Dose Limit: 125 PF 30PPS Dose Limit: 125 PPS Default: 15 PPS --------------------DAP Setup-------------------- DAP : off DAP Mode: summation DAP Device Type 1: 0 DAP Calibration Value 1: 1 DAP Test Value 1: 250 DAP Device Type 2: 0 DAP Calibration Value 2: 1 DAP Test Value 2: 250 DAP Accumulated Alarm 0 mGycm² DAP Rate Alarm 0 mGycm²/s DAP Printer off DAP Rate Display on