Kodak Dryview 5800 User Manual page 31

Laser imager
Hide thumbs Also See for Dryview 5800:
Table of Contents

Advertisement

Code
10003
MIS: Image Buffer Error
10015
MIS: Database Error
10910
MIS: MCS Communication
Failure
20004
USB Failure
20154
MCS: Internal Communications
Failure
20155
Incompatible MCS Printer
Configuration for hardware
20156
Incompatible Software Versions
Installed
20209
Laser Imager Opened During
Self Test
20449
none
20701
none
20702
none
2011-09-09
Table 3-5: Condition Codes (Continued)
Web Portal Message
User Action
1. Shut down the imager, then start the imager.
2. If the error persists, call for service.
1. Shut down the imager, then start the imager.
2. If the error persists, call for service.
1. Shut down the imager, then start the imager.
2. If the error persists, call for service.
1. Shut down the imager, then start the imager.
2. If the error persists, call for service.
1. Shut down the imager, then start the imager.
2. If the error persists, call for service.
1. Shut down the imager, then start the imager.
2. If the error persists, call for service.
1. Shut down the imager, then start the imager.
2. If the error persists, call for service.
1. Close the hood or the open cover.
2. Shut down the imager, then start the imager.
3. If the error persists, call for service.
1. Change the charcoal filter.
2. Press and hold the Menu Selection button for five
seconds. Release the button and the Test Print
menu displays.
3. Press and release the Menu Selection button again
to display the Error Reset menu.
4. Check that the 20449 error is highlighted on the
display screen, and press the Enter button.
The icon and condition code are cleared, and the
display screen returns to normal operation.
1. Close the hood or the open cover.
1. Install the missing film tray.
2G0733
Troubleshooting
3-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dryview 5850Carestream dryview 5850

Table of Contents