7.3.4 Error message: Water pump error

7.3.5 Error message: No chamb water

Revision: 2.1
Error message: Water pump error
No water detected at the high electrode
of the generator after approx. 2 minutes
(parameter 70,
"GenWtrErrT")
- Defective electrode
- Check the steam generator
(see chapter 3.8.1, "Steam
Error message: No chamb water
No water has been measured at the water
level electrode (E3) of the sterilisation
chamber approx. 5 minutes
(parameter 94,
"WtrChambErr")
after the program starts
- Check function and cleanliness of
water inlet valve
- Check and clean the strainer
- Check the water level electrode
V/D S
T
ROUBLESHOOTING
generator")
ERIES
7-7