Filter Parameters: Date & Time : Last 07 Days Domain : application, system, security, customerlog Source : * Severity : warning, error Facility : Service MessageText : MessageId : Search Order : Newest first No. of occurances: 1000 ------------ S. Date/Time ID Source Seq. Message Text -- --------- -- ------ ---- ------------ E 14-Oct-2014 18:26:15 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 18:26:15 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 18:26:15 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 18:26:15 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 18:26:15 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 18:26:15 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 18:26:14 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:25:41 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 18:24:31 22 MRI_MSR 0 Start of Measurement System failed. E 14-Oct-2014 18:24:31 22 MRI_MSR 4 Start of Measurement System failed. E 14-Oct-2014 18:24:31 23 MRI_MSR 3 Start of Scanner failed. E 14-Oct-2014 18:24:31 5 MRI_SUP 2 Scanner hardware error: A peripheral unit of the Scanner is not ready for measurement. Check other error messages for details. Reboot scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:24:31 5 MRI_SUP 1 Scanner hardware error: A peripheral unit of the Scanner is not ready for measurement. Check other error messages for details. Reboot scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:24:31 5 MRI_SUP 0 Scanner hardware error: A peripheral unit of the Scanner is not ready for measurement. Check other error messages for details. Reboot scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:24:31 5 MRI_SUP 0 Scanner hardware error: A peripheral unit of the Scanner is not ready for measurement. Check other error messages for details. Reboot scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:24:31 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 18:24:29 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 18:24:10 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 18:20:51 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 18:20:08 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 18:20:08 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:16:27 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:16:27 36 MRI_PER 0 Booting the Scanner has failed. See previous messages for details. E 14-Oct-2014 18:16:27 1 CAP_DVF 0 An unspecified internal error has occurred. W 14-Oct-2014 18:16:19 14 MRI_PIA 0 No peakmeter available for line WaveIn. W 14-Oct-2014 18:16:19 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 18:16:15 10 MRI_SC 0 Sequence control error: Failed to send a DSP message. Retry, reboot Scanner. E 14-Oct-2014 18:16:15 10 MRI_SC 0 Sequence control error: Failed to send a DSP message. Retry, reboot Scanner. E 14-Oct-2014 18:16:01 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 18:15:58 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 18:15:58 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 18:15:49 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection. E 14-Oct-2014 18:15:47 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 18:11:56 14 Hummingbird NFS Maestro Server n/a Can not unregister program '100005' with Portmapper. Error=10054 An existing connection was forcibly closed by the remote host. W 14-Oct-2014 18:11:56 1517 Userenv n/a Windows saved user MRC27010\meduser registry while an application or service was still using the registry during log off. The memory used by the user's registry has not been freed. The registry will be unloaded when it is no longer in use. This is often caused by services running as a user account, try configuring the services to run in either the LocalService or NetworkService account. E 14-Oct-2014 18:11:17 11 MRI_COS 6 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 18:11:17 11 MRI_COS 5 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 18:11:17 11 MRI_COS 3 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 18:11:17 11 MRI_COS 2 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 18:11:17 32 MRI_COS 1 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 18:11:17 32 MRI_COS 0 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 18:11:06 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details. E 14-Oct-2014 18:11:04 1 MRI_PER 2 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 18:11:04 1 MRI_PER 1 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 18:11:04 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 18:10:47 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 18:10:44 1 MRI_PER 30 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 18:10:44 55 MRI_PMU 28 Physiological Measurement Unit: PERU receive data warning. W 14-Oct-2014 18:10:44 54 MRI_PMU 27 Physiological Measurement Unit: PPU: Pulse Finger clip not attached properly. E 14-Oct-2014 18:10:44 661 MRI_SUP 26 Tx DSP error: No communication with unit Tx DSP. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:10:44 37 MRI_PXX 25 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 18:10:44 37 MRI_PXX 24 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 18:10:44 3000 MRI_MSP 23 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 430 MRI_SUP 22 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 440 MRI_SUP 21 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 3000 MRI_PXX 20 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:10:44 36 MRI_TXD 19 RF control error: PCITX is still in test mode. Reboot scanner in order to leave testmode. E 14-Oct-2014 18:10:44 27 MRI_SLO 13 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 30 MRI_SLO 12 RF Power Amplifier error: RFPA power supply current too high. Retry, check sequence and/or sequence parameter If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 29 MRI_SLO 11 RF Power Amplifier error: RF pulse too long or pulse time interval too short. Retry, check sequence and/or sequence parameter. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 28 MRI_SLO 10 RF Power Amplifier error: Temperature of the RFPA too high. Retry after 5 minutes. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 27 MRI_SLO 9 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 9 MRI_SUP 8 RF Power Amplifier error: RFPA unit was switched off during measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 2001 MRI_GXX 7 Gradient Power Amplifier error: While scanning, the power supply for the GPA was switched off. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:10:44 2000 MRI_GXX 6 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:10:44 28 MRI_PTX 5 GC FIFO underflow If the problem persists: please, contact service with the hint that details are available in service level. E 14-Oct-2014 18:10:44 27 MRI_SLO 4 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 30 MRI_SLO 3 RF Power Amplifier error: RFPA power supply current too high. Retry, check sequence and/or sequence parameter If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 29 MRI_SLO 2 RF Power Amplifier error: RF pulse too long or pulse time interval too short. Retry, check sequence and/or sequence parameter. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:44 28 MRI_SLO 1 RF Power Amplifier error: Temperature of the RFPA too high. Retry after 5 minutes. If the problem still persists, please call Siemens Service. W 14-Oct-2014 18:10:39 55 MRI_PMU 30 Physiological Measurement Unit: PERU receive data warning. W 14-Oct-2014 18:10:39 54 MRI_PMU 29 Physiological Measurement Unit: PPU: Pulse Finger clip not attached properly. E 14-Oct-2014 18:10:39 661 MRI_SUP 28 Tx DSP error: No communication with unit Tx DSP. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:10:39 37 MRI_PXX 27 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 18:10:39 37 MRI_PXX 26 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 18:10:39 3000 MRI_MSP 25 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 430 MRI_SUP 24 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 440 MRI_SUP 23 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 3000 MRI_PXX 22 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:10:39 36 MRI_TXD 21 RF control error: PCITX is still in test mode. Reboot scanner in order to leave testmode. E 14-Oct-2014 18:10:39 27 MRI_SLO 15 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 30 MRI_SLO 14 RF Power Amplifier error: RFPA power supply current too high. Retry, check sequence and/or sequence parameter If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 29 MRI_SLO 13 RF Power Amplifier error: RF pulse too long or pulse time interval too short. Retry, check sequence and/or sequence parameter. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 28 MRI_SLO 12 RF Power Amplifier error: Temperature of the RFPA too high. Retry after 5 minutes. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 27 MRI_SLO 11 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 9 MRI_SUP 10 RF Power Amplifier error: RFPA unit was switched off during measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 2001 MRI_GXX 9 Gradient Power Amplifier error: While scanning, the power supply for the GPA was switched off. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:10:39 2000 MRI_GXX 8 Gradient Power Amplifier error: The GPA is not ready for measurement. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:10:39 28 MRI_PTX 7 GC FIFO underflow If the problem persists: please, contact service with the hint that details are available in service level. E 14-Oct-2014 18:10:39 27 MRI_SLO 6 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 30 MRI_SLO 5 RF Power Amplifier error: RFPA power supply current too high. Retry, check sequence and/or sequence parameter If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 29 MRI_SLO 4 RF Power Amplifier error: RF pulse too long or pulse time interval too short. Retry, check sequence and/or sequence parameter. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 28 MRI_SLO 3 RF Power Amplifier error: Temperature of the RFPA too high. Retry after 5 minutes. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 27 MRI_SLO 2 RF Power Amplifier error: Power failure message from RFPA. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:10:39 27 MRI_PAD 1 RF Power Absorption Limiter error: RF limit for 5th averaging stage exceeded. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 18:07:27 21 MRI_MSR 0 Preparation of Measurement System failed. E 14-Oct-2014 18:07:27 21 MRI_MSR 1 Preparation of Measurement System failed. E 14-Oct-2014 18:07:27 56 MRI_RFS 0 RF Safety Watchdog error: Internal software error (invalid parameter supplied to a RFSWD method). Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 18:07:27 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. W 14-Oct-2014 18:07:22 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 18:07:22 56 MRI_RFS 3 RF Safety Watchdog error: Internal software error (invalid parameter supplied to a RFSWD method). Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 18:07:22 56 MRI_RFS 2 RF Safety Watchdog error: Internal software error (invalid parameter supplied to a RFSWD method). Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 18:07:22 56 MRI_RFS 1 RF Safety Watchdog error: Internal software error (invalid parameter supplied to a RFSWD method). Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 18:07:22 56 MRI_RFS 0 RF Safety Watchdog error: Internal software error (invalid parameter supplied to a RFSWD method). Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 18:07:22 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:07:22 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:07:22 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:22 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:17 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:07:17 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:07:17 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:17 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:11 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:07:11 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:07:11 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:11 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:06 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:07:06 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:07:06 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:06 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:00 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:07:00 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:07:00 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:07:00 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:55 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:55 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:55 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:55 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:49 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:49 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:49 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:49 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:44 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:44 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:44 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:44 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:39 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:39 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:39 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:39 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:33 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:33 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:33 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:33 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:28 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:28 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:28 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:28 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:22 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:22 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:22 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:22 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:17 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:17 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:17 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:17 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:11 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:11 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:11 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:11 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:06 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:06 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:06 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:06 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:01 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:06:01 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:06:01 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:06:01 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:55 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:55 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:55 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:55 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:50 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:50 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:50 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:50 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:44 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:44 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:44 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:44 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:39 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:39 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:39 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:39 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:34 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:34 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:34 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:34 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:28 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:28 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:28 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:28 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:23 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:23 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:23 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:23 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:17 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:17 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:17 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:17 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:12 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:12 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:12 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:12 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:06 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:06 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:06 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:06 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:01 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:05:01 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:05:01 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:05:01 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:56 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:56 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:56 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:55 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:50 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:50 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:50 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:50 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:45 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:45 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:45 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:45 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:39 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:39 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:39 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:39 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:34 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:34 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:34 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:34 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:28 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:28 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:28 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:28 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:23 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:23 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:23 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:23 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:18 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:18 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:18 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:18 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:12 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:12 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:12 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:12 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:07 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:07 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:07 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:07 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:01 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:04:01 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:04:01 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:04:01 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:56 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:56 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:56 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:56 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:51 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:51 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:51 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:51 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:45 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:45 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:45 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:45 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:40 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:40 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:40 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:40 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:34 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:34 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:34 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:34 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:29 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:29 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:29 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:29 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:24 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:24 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:24 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:24 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:18 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:18 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:18 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:18 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:13 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:13 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:13 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:13 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:07 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:07 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:07 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:07 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:02 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:03:02 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:03:02 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:03:02 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:57 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:02:56 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:02:56 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:56 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:51 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:02:51 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:02:51 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:51 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:46 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:02:46 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:02:46 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:46 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:40 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:02:40 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:02:40 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:40 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:35 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:02:35 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:02:35 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:35 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:30 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:02:30 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:02:30 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:30 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:24 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:02:24 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:02:24 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:24 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:07 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:02:07 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:02:07 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:02:07 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:57 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:01:57 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:01:57 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:57 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:47 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:01:47 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:01:47 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:47 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:37 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:01:37 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:01:37 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:37 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:27 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:01:27 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:01:27 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:27 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:17 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:01:17 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:01:17 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:17 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:12 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:01:12 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:01:12 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:12 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:06 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:01:06 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:01:06 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:06 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:01 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:01:01 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:01:01 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:01:01 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:55 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:00:55 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:00:55 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:55 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:50 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:00:50 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:00:50 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:50 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:45 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:00:45 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:00:45 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:45 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:39 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:00:39 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:00:39 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:39 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:34 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:00:34 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:00:34 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:34 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:28 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:00:28 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:00:28 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:28 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:04 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 18:00:04 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 18:00:04 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 18:00:04 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:59:55 19 MRI_MSR 0 Preparation of Scanner failed. E 14-Oct-2014 17:59:55 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 17:59:55 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:59:55 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:54:34 2010 MRI_ADS 0 Frequency adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 17:54:34 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 17:54:34 19 MRI_MSR 1 Preparation of Scanner failed. E 14-Oct-2014 17:54:34 46 MRI_MSR 0 Could not prepare sequence on scanner Retry, reboot Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:54:34 999 MRI_SSL 0 Sequence run-time error: Inernal error occurred. Retry, reboot scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:54:34 661 MRI_SUP 0 Tx DSP error: No communication with unit Tx DSP. Retry, reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service E 14-Oct-2014 17:54:02 2010 MRI_ADS 0 Frequency adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 17:54:02 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 17:54:02 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 17:54:02 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 17:54:02 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 17:54:02 3000 MRI_MSP 0 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:54:02 430 MRI_SUP 0 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:54:02 440 MRI_SUP 0 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:54:02 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 17:53:48 36 MRI_TXD 0 RF control error: PCITX is still in test mode. Reboot scanner in order to leave testmode. W 14-Oct-2014 17:53:26 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 17:46:56 28 MRI_PTX 0 GC FIFO underflow If the problem persists: please, contact service with the hint that details are available in service level. E 14-Oct-2014 17:45:39 29 MRI_SLO 0 RF Power Amplifier error: RF pulse too long or pulse time interval too short. Retry, check sequence and/or sequence parameter. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:45:05 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 17:44:21 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 17:44:21 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:28:00 28 MRI_PTX 0 GC FIFO underflow If the problem persists: please, contact service with the hint that details are available in service level. E 14-Oct-2014 17:23:34 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 17:22:52 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 17:22:52 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:22:13 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 17:22:13 36 MRI_PER 0 Booting the Scanner has failed. See previous messages for details. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:22:13 13 MRI_MCO 0 Scanner communication error: Data transfer to Scanner failed. Retry, reboot Scanner, switch off/on Scanner, reboot syngo MR. E 14-Oct-2014 17:21:50 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 17:21:01 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 17:21:01 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:57:42 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 16:57:42 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 16:57:42 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 16:57:42 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 16:57:42 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 16:57:42 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:57:42 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 16:57:09 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 16:57:09 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 16:57:09 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 16:57:09 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 16:57:09 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 16:57:09 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:57:09 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 16:56:43 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 16:46:38 1 CAP_DVF 0 An unspecified internal error has occurred. W 14-Oct-2014 16:46:30 14 MRI_PIA 0 No peakmeter available for line WaveIn. W 14-Oct-2014 16:46:30 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 16:46:26 40 MRI_PER 0 Patient table error: Patient table position could not be determined. Move patient table to home position. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any message window, before the function has been finished successfully. Please call Siemens Service E 14-Oct-2014 16:46:14 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 16:46:11 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 16:46:10 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 16:46:01 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection. E 14-Oct-2014 16:46:00 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 16:28:02 1517 Userenv n/a Windows saved user MRC27010\meduser registry while an application or service was still using the registry during log off. The memory used by the user's registry has not been freed. The registry will be unloaded when it is no longer in use. This is often caused by services running as a user account, try configuring the services to run in either the LocalService or NetworkService account. E 14-Oct-2014 16:27:36 32 MRI_COS 6 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 16:27:36 11 MRI_COS 5 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 16:27:36 11 MRI_COS 3 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 16:27:36 11 MRI_COS 2 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 16:27:36 32 MRI_COS 1 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 16:27:36 32 MRI_COS 0 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 16:27:25 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details. E 14-Oct-2014 16:27:24 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 16:27:03 1 MRI_PER 30 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 16:27:03 200 MRI_SUP 28 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 16:27:03 801 MRI_COH 27 Scanner control error: Measurement is blocked by the MSUP supervisory test. Retry after the supervisory test is finished. E 14-Oct-2014 16:27:03 67 MRI_MSP 25 Magnet Supervision message: The MSU selftest is active, no scanning possible at this time. Wait for end of MSU selftest (approx. 1 minute). If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:27:03 801 MRI_COH 24 Scanner control error: Measurement is blocked by the MSUP supervisory test. Retry after the supervisory test is finished. W 14-Oct-2014 16:27:03 56 MRI_PMU 22 Physiological Measurement Unit: Rechargeable PERU battery weak. E 14-Oct-2014 16:27:03 37 MRI_PXX 21 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:27:03 37 MRI_PXX 20 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:27:03 3000 MRI_PXX 19 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 16:27:03 43 MRI_SUP 13 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:27:03 43 MRI_SUP 12 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:27:03 3000 MRI_MSP 6 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:27:03 3000 MRI_PXX 5 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 16:27:03 440 MRI_SUP 4 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:27:03 430 MRI_SUP 3 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:27:03 43 MRI_SUP 2 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:27:03 43 MRI_SUP 1 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:26:58 200 MRI_SUP 30 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 16:26:58 801 MRI_COH 29 Scanner control error: Measurement is blocked by the MSUP supervisory test. Retry after the supervisory test is finished. E 14-Oct-2014 16:26:58 67 MRI_MSP 27 Magnet Supervision message: The MSU selftest is active, no scanning possible at this time. Wait for end of MSU selftest (approx. 1 minute). If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:26:58 801 MRI_COH 26 Scanner control error: Measurement is blocked by the MSUP supervisory test. Retry after the supervisory test is finished. W 14-Oct-2014 16:26:58 56 MRI_PMU 24 Physiological Measurement Unit: Rechargeable PERU battery weak. E 14-Oct-2014 16:26:58 37 MRI_PXX 23 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:26:58 37 MRI_PXX 22 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:26:58 3000 MRI_PXX 21 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 16:26:58 43 MRI_SUP 15 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:26:58 43 MRI_SUP 14 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:26:58 3000 MRI_MSP 8 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:26:58 3000 MRI_PXX 7 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 16:26:58 440 MRI_SUP 6 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:26:58 430 MRI_SUP 5 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:26:58 43 MRI_SUP 4 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:26:58 43 MRI_SUP 3 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:26:58 37 MRI_PXX 2 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:26:58 37 MRI_PXX 1 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 16:26:18 200 MRI_SUP 0 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 16:13:24 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 16:13:24 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 16:13:24 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 16:13:24 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 16:13:24 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 16:13:24 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:13:23 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 16:13:06 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 16:12:35 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 16:09:48 1 CAP_DVF 0 An unspecified internal error has occurred. W 14-Oct-2014 16:09:40 14 MRI_PIA 0 No peakmeter available for line WaveIn. W 14-Oct-2014 16:09:39 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 16:09:23 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 16:09:20 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 16:09:19 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 16:09:10 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection. E 14-Oct-2014 16:09:09 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 16:02:07 1517 Userenv n/a Windows saved user MRC27010\meduser registry while an application or service was still using the registry during log off. The memory used by the user's registry has not been freed. The registry will be unloaded when it is no longer in use. This is often caused by services running as a user account, try configuring the services to run in either the LocalService or NetworkService account. E 14-Oct-2014 16:01:44 32 MRI_COS 4 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 16:01:44 11 MRI_COS 3 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 16:01:44 11 MRI_COS 1 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 16:01:44 32 MRI_COS 0 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 16:01:33 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details. E 14-Oct-2014 16:01:32 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 16:01:02 1 MRI_PER 30 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 16:01:02 200 MRI_SUP 28 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 16:01:02 37 MRI_PXX 27 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:01:02 37 MRI_PXX 26 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:01:02 3000 MRI_PXX 25 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 16:01:02 3000 MRI_MSP 19 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:01:02 3000 MRI_PXX 18 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 16:01:02 440 MRI_SUP 17 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:01:02 430 MRI_SUP 16 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:01:02 43 MRI_SUP 15 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:01:02 43 MRI_SUP 14 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:01:02 43 MRI_SUP 8 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:01:02 43 MRI_SUP 7 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:01:02 43 MRI_SUP 1 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:00:59 200 MRI_SUP 30 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 16:00:59 37 MRI_PXX 29 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:00:59 37 MRI_PXX 28 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 16:00:59 3000 MRI_PXX 27 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 16:00:59 3000 MRI_MSP 21 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:00:59 3000 MRI_PXX 20 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 16:00:59 440 MRI_SUP 19 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 16:00:59 430 MRI_SUP 18 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:00:59 43 MRI_SUP 17 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:00:59 43 MRI_SUP 16 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:00:59 43 MRI_SUP 10 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:00:59 43 MRI_SUP 9 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:00:59 43 MRI_SUP 3 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:00:59 43 MRI_SUP 2 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 16:00:27 200 MRI_SUP 0 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 15:59:31 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 15:59:31 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 15:59:31 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 15:59:31 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 15:59:31 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 15:59:31 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 15:59:31 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 15:59:02 14 MRI_PIA 0 No peakmeter available for line WaveIn. W 14-Oct-2014 15:34:57 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 15:33:37 1 CAP_DVF 0 An unspecified internal error has occurred. W 14-Oct-2014 15:33:28 14 MRI_PIA 0 No peakmeter available for line WaveIn. W 14-Oct-2014 15:33:28 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 15:33:13 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 15:33:10 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 15:33:10 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 15:33:01 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection. E 14-Oct-2014 15:33:00 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 15:28:10 11 MRI_COS 6 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 15:28:10 11 MRI_COS 5 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 15:28:10 11 MRI_COS 3 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 15:28:10 11 MRI_COS 2 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 15:28:10 32 MRI_COS 1 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 15:28:10 32 MRI_COS 0 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 15:27:59 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details. E 14-Oct-2014 15:27:57 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 15:27:27 1 MRI_PER 30 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 15:27:27 52 MRI_PMU 28 Physiological Measurement Unit: PPU data receive warning. E 14-Oct-2014 15:27:27 511 MRI_SUP 26 Scanner control error: Coil configuration change was detected while scanning. Check coil plugs for proper connection. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:27 511 MRI_SUP 24 Scanner control error: Coil configuration change was detected while scanning. Check coil plugs for proper connection. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:27 511 MRI_SUP 22 Scanner control error: Coil configuration change was detected while scanning. Check coil plugs for proper connection. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:27 511 MRI_SUP 20 Scanner control error: Coil configuration change was detected while scanning. Check coil plugs for proper connection. If the problem still persists, please call Siemens Service. W 14-Oct-2014 15:27:27 52 MRI_PMU 19 Physiological Measurement Unit: PPU data receive warning. W 14-Oct-2014 15:27:27 54 MRI_PMU 18 Physiological Measurement Unit: PPU: Pulse Finger clip not attached properly. W 14-Oct-2014 15:27:27 53 MRI_PMU 17 Physiological Measurement Unit: Rechargeable PPU battery weak. W 14-Oct-2014 15:27:27 52 MRI_PMU 16 Physiological Measurement Unit: PPU data receive warning. W 14-Oct-2014 15:27:27 54 MRI_PMU 15 Physiological Measurement Unit: PPU: Pulse Finger clip not attached properly. W 14-Oct-2014 15:27:27 53 MRI_PMU 14 Physiological Measurement Unit: Rechargeable PPU battery weak. W 14-Oct-2014 15:27:27 52 MRI_PMU 13 Physiological Measurement Unit: PPU data receive warning. E 14-Oct-2014 15:27:27 37 MRI_PXX 12 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 15:27:27 37 MRI_PXX 11 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 15:27:27 3000 MRI_PXX 10 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 15:27:27 3000 MRI_MSP 4 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:27 3000 MRI_PXX 3 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 15:27:27 440 MRI_SUP 2 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:27 430 MRI_SUP 1 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 15:27:24 52 MRI_PMU 30 Physiological Measurement Unit: PPU data receive warning. E 14-Oct-2014 15:27:24 511 MRI_SUP 28 Scanner control error: Coil configuration change was detected while scanning. Check coil plugs for proper connection. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:24 511 MRI_SUP 26 Scanner control error: Coil configuration change was detected while scanning. Check coil plugs for proper connection. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:24 511 MRI_SUP 24 Scanner control error: Coil configuration change was detected while scanning. Check coil plugs for proper connection. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:24 511 MRI_SUP 22 Scanner control error: Coil configuration change was detected while scanning. Check coil plugs for proper connection. If the problem still persists, please call Siemens Service. W 14-Oct-2014 15:27:24 52 MRI_PMU 21 Physiological Measurement Unit: PPU data receive warning. W 14-Oct-2014 15:27:24 54 MRI_PMU 20 Physiological Measurement Unit: PPU: Pulse Finger clip not attached properly. W 14-Oct-2014 15:27:24 53 MRI_PMU 19 Physiological Measurement Unit: Rechargeable PPU battery weak. W 14-Oct-2014 15:27:24 52 MRI_PMU 18 Physiological Measurement Unit: PPU data receive warning. W 14-Oct-2014 15:27:24 54 MRI_PMU 17 Physiological Measurement Unit: PPU: Pulse Finger clip not attached properly. W 14-Oct-2014 15:27:24 53 MRI_PMU 16 Physiological Measurement Unit: Rechargeable PPU battery weak. W 14-Oct-2014 15:27:24 52 MRI_PMU 15 Physiological Measurement Unit: PPU data receive warning. E 14-Oct-2014 15:27:24 37 MRI_PXX 14 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 15:27:24 37 MRI_PXX 13 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 15:27:24 3000 MRI_PXX 12 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 15:27:24 3000 MRI_MSP 6 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:24 3000 MRI_PXX 5 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 15:27:24 440 MRI_SUP 4 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:27:24 430 MRI_SUP 3 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 15:27:24 43 MRI_SUP 2 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 15:27:24 43 MRI_SUP 1 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 15:18:05 4 MRI_CCS 0 Coil configuration error: Coil "Spine Matrix" with more than one plug has been connected incompletely. Make sure all coil conectors are connected properly. E 14-Oct-2014 15:15:38 4 MRI_CCS 0 Coil configuration error: Coil "Spine Matrix" with more than one plug has been connected incompletely. Make sure all coil conectors are connected properly. W 14-Oct-2014 14:35:55 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 14:35:18 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 14:35:18 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 14:35:18 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 14:35:18 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:35:18 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:35:17 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:35:17 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:34:53 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 14:34:53 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 14:34:53 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 14:34:53 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:34:53 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:34:53 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:34:52 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:34:32 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 14:34:32 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 14:34:32 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 14:34:32 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:34:32 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:34:32 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:34:31 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:34:16 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 14:34:16 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:34:16 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 14:34:16 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:34:16 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:34:16 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:34:15 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:33:29 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 14:33:29 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 14:33:29 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 14:33:29 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:33:29 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:33:29 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:33:29 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:32:55 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 14:32:55 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:32:55 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 14:32:55 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:32:55 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:32:55 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:32:54 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:31:25 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 14:31:25 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 14:31:25 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 14:31:25 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:31:25 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:31:25 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:31:24 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 14:30:43 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 14:28:23 1 CAP_DVF 0 An unspecified internal error has occurred. W 14-Oct-2014 14:28:15 14 MRI_PIA 0 No peakmeter available for line WaveIn. W 14-Oct-2014 14:28:15 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 14:27:59 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 14:27:56 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 14:27:55 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 14:27:47 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection. E 14-Oct-2014 14:27:45 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 14:19:43 1517 Userenv n/a Windows saved user MRC27010\meduser registry while an application or service was still using the registry during log off. The memory used by the user's registry has not been freed. The registry will be unloaded when it is no longer in use. This is often caused by services running as a user account, try configuring the services to run in either the LocalService or NetworkService account. E 14-Oct-2014 14:19:21 32 MRI_COS 6 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 14:19:21 11 MRI_COS 5 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 14:19:21 11 MRI_COS 3 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 14:19:21 11 MRI_COS 2 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 14:19:21 32 MRI_COS 1 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 14:19:21 32 MRI_COS 0 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 14:19:11 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details. E 14-Oct-2014 14:19:09 1 MRI_PER 1 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 14:19:09 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 14:18:21 1 MRI_PER 30 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 14:18:21 200 MRI_SUP 28 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. W 14-Oct-2014 14:18:21 54 MRI_PMU 27 Physiological Measurement Unit: PPU: Pulse Finger clip not attached properly. E 14-Oct-2014 14:18:21 37 MRI_PXX 26 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:18:21 37 MRI_PXX 25 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:18:21 3000 MRI_PXX 24 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 14:18:21 43 MRI_SUP 18 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 14:18:21 43 MRI_SUP 17 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 14:18:21 3000 MRI_MSP 11 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 14:18:21 3000 MRI_PXX 10 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:18:21 440 MRI_SUP 9 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 14:18:21 430 MRI_SUP 8 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 14:18:21 43 MRI_SUP 7 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 14:18:21 43 MRI_SUP 6 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 14:18:21 37 MRI_PXX 5 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:18:21 37 MRI_PXX 4 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 14:18:18 200 MRI_SUP 30 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. W 14-Oct-2014 14:18:18 54 MRI_PMU 29 Physiological Measurement Unit: PPU: Pulse Finger clip not attached properly. E 14-Oct-2014 14:18:18 37 MRI_PXX 28 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:18:18 37 MRI_PXX 27 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:18:18 3000 MRI_PXX 26 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 14:18:18 43 MRI_SUP 20 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 14:18:18 43 MRI_SUP 19 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 14:18:18 3000 MRI_MSP 13 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 14:18:18 3000 MRI_PXX 12 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:18:18 440 MRI_SUP 11 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 14:18:18 430 MRI_SUP 10 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 14:18:18 43 MRI_SUP 9 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 14:18:18 43 MRI_SUP 8 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 14:18:18 37 MRI_PXX 7 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:18:18 37 MRI_PXX 6 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 14:17:44 200 MRI_SUP 0 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 14:13:58 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 14:13:58 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 14:13:58 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 14:13:58 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:13:58 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:13:58 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:13:57 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:13:35 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:11:45 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 14:11:04 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 14:11:04 15 MRI_PER 0 Scanner is not ready for measurement. Reboot Scanner, switch off/on Scanner. If the problem still persists, please call Siemens Service. E 14-Oct-2014 14:09:28 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 14:09:28 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:09:28 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 14:09:28 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:09:28 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:09:28 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:09:28 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:04:26 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 14:04:26 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:04:26 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 14:04:26 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:04:26 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:04:26 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:04:26 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:03:58 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 14:03:58 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:03:58 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 14:03:58 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 14:03:58 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 14:03:58 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 14:03:58 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 14:03:01 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 14:02:25 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 13:58:33 1 CAP_DVF 0 An unspecified internal error has occurred. W 14-Oct-2014 13:58:25 14 MRI_PIA 0 No peakmeter available for line WaveIn. W 14-Oct-2014 13:58:25 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 13:58:08 34 MRI_PER 0 Scanner is booting. E 14-Oct-2014 13:58:05 35 MRI_PER 0 Scanner is not online (might be switched off). E 14-Oct-2014 13:58:05 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 13:57:56 101 MRI_ISC 0 Magellan Mouse Driver error: Internal software error (Create file error). Check 6d mouse configuration/connection. E 14-Oct-2014 13:57:54 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 13:37:26 1517 Userenv n/a Windows saved user MRC27010\meduser registry while an application or service was still using the registry during log off. The memory used by the user's registry has not been freed. The registry will be unloaded when it is no longer in use. This is often caused by services running as a user account, try configuring the services to run in either the LocalService or NetworkService account. E 14-Oct-2014 13:37:00 32 MRI_COS 8 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 13:37:00 11 MRI_COS 7 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 13:37:00 11 MRI_COS 5 The image reconstruction system is not ready. If the system is rebooting at the moment please wait for system startup. Otherwise reboot image reconstruction system. If this is not possible switch to Standby and turn on Scanner again afterwords. E 14-Oct-2014 13:37:00 32 MRI_COS 4 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 13:37:00 32 MRI_COS 3 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 13:37:00 32 MRI_COS 2 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 13:37:00 32 MRI_COS 1 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 13:37:00 32 MRI_COS 0 Connection to Image Reconstruction System has been lost. E 14-Oct-2014 13:36:49 45 MRI_OLD 1 An error occurred during accessing db. See logfile for details. E 14-Oct-2014 13:36:48 1 MRI_PER 0 Scanner is not online. Wait until booting has finished, or switch on Scanner. E 14-Oct-2014 13:36:18 1 MRI_PER 30 Scanner is not online. Wait until booting has finished, or switch on Scanner. W 14-Oct-2014 13:36:18 200 MRI_SUP 28 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 13:36:18 37 MRI_PXX 27 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:36:18 37 MRI_PXX 26 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:36:18 3000 MRI_PXX 25 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 13:36:18 43 MRI_SUP 19 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 13:36:18 43 MRI_SUP 18 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 13:36:18 3000 MRI_MSP 12 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 13:36:18 3000 MRI_PXX 11 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:36:18 440 MRI_SUP 10 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 13:36:18 430 MRI_SUP 9 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 13:36:18 43 MRI_SUP 8 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 13:36:18 43 MRI_SUP 7 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 13:36:18 37 MRI_PXX 6 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:36:18 37 MRI_PXX 5 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:36:18 3000 MRI_PXX 4 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 13:36:14 200 MRI_SUP 30 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 13:36:14 37 MRI_PXX 29 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:36:14 37 MRI_PXX 28 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:36:14 3000 MRI_PXX 27 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 13:36:14 43 MRI_SUP 21 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 13:36:14 43 MRI_SUP 20 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 13:36:14 3000 MRI_MSP 14 MSUP warning: While scanning, the Magnet Supervision left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 13:36:14 3000 MRI_PXX 13 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:36:14 440 MRI_SUP 12 RF-Infra structure error: While scanning, the RF-Infra structure left the "sleep" mode because of a coil configuration change, a powerfail or wake up by other components. This could cause noise in the images. Check coil plugs for proper connection. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service. If the problem still persists, please call Siemens Service. E 14-Oct-2014 13:36:14 430 MRI_SUP 11 Receive Channel Selector error: While scanning, the Receive Channel Selector left the "sleep" mode because of Powerfail or wake up by other components. This could cause noise in the images. Start sequence again. If not successful, reboot Scanner to solve a possible software problem. If not successful, please call Siemens Service If the problem still persists, please call Siemens Service. W 14-Oct-2014 13:36:14 43 MRI_SUP 10 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. W 14-Oct-2014 13:36:14 43 MRI_SUP 9 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 13:36:14 37 MRI_PXX 8 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:36:14 37 MRI_PXX 7 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:36:14 3000 MRI_PXX 6 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 13:35:40 200 MRI_SUP 0 Scanner hardware warning: Failure of the cooling fans for the Imager 1. Switch off/on Scanner. Please call Siemens Service. E 14-Oct-2014 13:33:00 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:33:00 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:33:00 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:33:00 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:33:00 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:33:00 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:32:59 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:31:58 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:31:58 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:31:58 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:31:58 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:31:58 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:31:58 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:31:57 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:28:40 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:28:40 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:28:40 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:28:40 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:28:40 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:28:40 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:28:40 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:27:14 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:27:14 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:27:14 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:27:14 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:27:14 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:27:14 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:27:13 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:25:54 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:25:54 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:25:54 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:25:54 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:25:54 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:25:54 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:25:53 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:24:55 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:24:55 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:24:55 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:24:55 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:24:55 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:24:55 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:24:55 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:23:42 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:23:42 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:23:42 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:23:42 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:23:42 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:23:41 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:23:41 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:22:15 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:22:15 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:22:15 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:22:15 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:22:15 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:22:14 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:22:14 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:21:30 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:20:34 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:20:34 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:20:34 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:20:34 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:20:34 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:20:34 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:20:33 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 13:20:26 43 MRI_SUP 0 RF Power Amplifier warning: While scanning, the RFPA left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service. E 14-Oct-2014 13:19:05 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:19:05 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:19:05 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:19:05 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:19:05 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:19:04 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:19:04 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:11:18 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:11:18 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:11:18 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:11:18 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:11:18 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:11:18 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:11:17 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 13:11:08 576 MRI_ACS 0 SEP warning: SEP primary water temperature out of limits. Please call Siemens Service. E 14-Oct-2014 13:10:55 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:10:55 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:10:55 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:10:55 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:10:55 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:10:55 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:10:55 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 13:10:42 576 MRI_ACS 0 SEP warning: SEP primary water temperature out of limits. Please call Siemens Service. W 14-Oct-2014 13:10:42 576 MRI_ACS 0 SEP warning: SEP primary water temperature out of limits. Please call Siemens Service. E 14-Oct-2014 13:10:30 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:10:30 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:10:30 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:10:30 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:10:30 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:10:30 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:10:29 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service W 14-Oct-2014 13:10:21 576 MRI_ACS 0 SEP warning: SEP primary water temperature out of limits. Please call Siemens Service. E 14-Oct-2014 13:10:10 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:10:10 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:10:10 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:10:10 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:10:10 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:10:10 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:10:09 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:09:47 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:09:47 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:09:47 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:09:47 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:09:47 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:09:47 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:09:46 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:09:21 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:09:21 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:09:21 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:09:21 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:09:21 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:09:21 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:09:20 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:08:54 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:08:54 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:08:54 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:08:54 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:08:54 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:08:54 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:08:54 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:08:29 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:08:29 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:08:29 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:08:29 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:08:29 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:08:29 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:08:29 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:07:55 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:07:55 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:07:55 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:07:55 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:07:55 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:07:55 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:07:55 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:07:33 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:07:33 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:07:33 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:07:33 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:07:33 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:07:33 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:07:32 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:06:11 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:06:11 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:06:11 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:06:11 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:06:11 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:06:11 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:06:11 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:05:51 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:05:51 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:05:51 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:05:51 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:05:51 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:05:51 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:05:51 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:05:30 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. W 14-Oct-2014 13:05:08 14 MRI_PIA 0 No peakmeter available for line WaveIn. E 14-Oct-2014 13:01:53 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:01:53 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:01:53 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:01:53 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:01:53 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:01:53 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:01:53 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:01:40 27 MRI_MIF 0 Preparation of measurement system failed. E 14-Oct-2014 13:01:40 2040 MRI_ADS 0 Fieldmap adjustment did not converge. System could not be adjusted. Retry, reboot system, call service. E 14-Oct-2014 13:01:40 1401 MRI_ADS 0 Adjustment could not be measured. Adjustment was aborted by measurement system. Retry, reboot system, call service. E 14-Oct-2014 13:01:40 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:01:40 24 MRI_MSR 0 Scanning stopped by Scanner. E 14-Oct-2014 13:01:39 37 MRI_PXX 0 Patient Table error: Communication Host fault Reboot Scanner, switch off/on Scanner. If the problem still persits, please save System Log-Files by selecting "System/Control/Tools" and "Save System Log-Files". Please call Siemens Service. E 14-Oct-2014 13:01:39 3000 MRI_PXX 0 Patient table error: While scanning, the Patient table left the "sleep" mode. This could cause noise in the images. Retry. If the problem still persists, please call Siemens Service E 14-Oct-2014 13:01:06 7 MRI_MIF 2 Image Calculation error. Retry, reboot syngo MR. If the problem still persists, please save System Log-Files by selecting System/Control/Tools. Do not close any pop-up menu manually before the function has been finished successfully. Please call Siemens Service. E 14-Oct-2014 13:01:06 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:01:06 12 MRI_MSR 0 Scanning stopped. E 14-Oct-2014 13:01:06 12 MRI_MSR 1 Scanning stopped. E 14-Oct-2014 13:01:06 24 MRI_MSR 0 Scanning stopped by Scanner.